Photovoltaic Development The Challenge of Measuring pH in Silver Nitrate Solutions

From Kajal Jaiswal, Mettler Toledo 3 min Reading Time

Related Vendors

Photovoltaics play an important role in the transition to renewable energies. Read this article to find out why the pH value of silver nitrate affects the quality of solar panels.

Impact of silver nitrate on the performance of solar panels(Source: ©  Awais - stock.adobe.com)
Impact of silver nitrate on the performance of solar panels
(Source: © Awais - stock.adobe.com)

As the demand for renewable energy grows rapidly, photovoltaic (PV) solar cells have become a cornerstone technology in sustainable power generation. Among the various materials used in the manufacturing of these cells, silver nitrate (AgNO3) plays a crucial role. It is used to produce silver paste, which forms the electrodes on the solar cell surface. These electrodes are crucial because they conduct the electricity generated by the cell. However, the efficiency and durability of solar cells depend heavily on maintaining precise control over the pH of the silver nitrate solution during the manufacturing process.

Why pH Matters in Solar Cells Manufacturing

The pH level of the silver nitrate solution influences the surface charge of silver particles in the paste. This charge affects how well the particles spread and stick to the solar cell’s surface. When the pH is properly balanced, the silver paste disperses evenly and adheres strongly, forming uniform electrodes. This uniformity is vital for creating a highly conductive path for electrons, which directly impacts the solar cell’s efficiency.

On the other hand, if the pH is too high or too low, the paste may clump or fail to bond properly. This can lead to uneven electrode thickness and weaker adhesion, causing electrical resistance and potentially reducing the cell’s lifespan.

Solution to Avoid Silver Chloride Precipitation

Measuring the pH of silver nitrate is not without its challenges. When silver nitrate solutions come into contact with potassium chloride (KCl), a common electrolyte used in many pH sensors, a white precipitate of silver chloride (AgCl) can form. This precipitate builds up on the sensor, clogging it and causing inaccurate readings or sensor failure. These complications make conventional pH measurement methods unreliable for this application.

This double reference electrolyte system enables you to obtain accurate and stable pH readings without the risk of sensor clogging, resulting in improved control over the manufacturing process.(Source:  Mettler Toledo)
This double reference electrolyte system enables you to obtain accurate and stable pH readings without the risk of sensor clogging, resulting in improved control over the manufacturing process.
(Source: Mettler Toledo)

To overcome these difficulties, Mettler Toledo offers the InLab Science Pro-ISM sensor, designed specifically for demanding applications like measuring the pH of silver nitrate. This sensor features a unique double reference electrolyte system with an outer exchangeable electrolyte chamber. The design allows for replacing the usual potassium chloride electrolyte (KCl) with potassium nitrate (KNO3), which prevents the formation of silver chloride precipitate.

Best Practices for Accurate pH Measurement

Getting reliable pH measurements requires more than just the right sensor. Here are some best practices to ensure consistent, precise results:

  • Sensor Selection: Choose the Inlab Science Pro-ISM sensor, as its design specifically addresses the problems caused by silver chloride precipitate.
  • Electrolyte Replacement: Switch the outer bridge electrolyte from KCl to KNO3. This simple change dramatically reduces precipitate buildup
  • Sensor Calibration Pre-soaking: Before each measurement, calibrate the sensor carefully. Soak it in the KNO3 electrolyte for at least 30 minutes prior to calibration to stabilize the readings.
  • Measurement Conditions: Always measure both pH buffers and samples at the same temperature to avoid variability.
  • Repeat Measurements: Take triplicate measurements for each sample to verify consistency
  • Sensor Maintenance: Regularly clean the sensor with a thiourea solution to remove any deposits and extend its lifespan.
  • Proper Storage: When not in use, store the sensor in a wetting cap filled with fresh KNO3 electrolyte to maintain its functionality
  • Verification and Recalibration: After every 10 samples, verify sensor accuracy and recalibrate if readings drift outside acceptable limits

By carefully controlling the pH of silver nitrate in the electrode production process, you can produce solar cells that are not only more efficient but also more durable. The electrodes formed with well-controlled pH provide excellent electrical conductivity and mechanical stability, meaning the solar cells perform better for longer periods.

This improved performance contributes directly to advancing renewable energy technology. More efficient solar cells mean more energy harvested from the sun, reducing reliance on fossil fuels and supporting a cleaner, sustainable future.

If you want to learn more, you can download the pH calibration poster or check out the detailed application note on measuring the pH of silver nitrate.

(ID:50650051)

Subscribe to the newsletter now

Don't Miss out on Our Best Content

By clicking on „Subscribe to Newsletter“ I agree to the processing and use of my data according to the consent form (please expand for details) and accept the Terms of Use. For more information, please see our Privacy Policy. The consent declaration relates, among other things, to the sending of editorial newsletters by email and to data matching for marketing purposes with selected advertising partners (e.g., LinkedIn, Google, Meta)